Free Call Australia : 1 800 834 991

Related Products

Hioki ST5540 Leak Current HiTester

HIOKI ST5540 Leak Current HiTester 
  • Uninterrupted polarity switching function dramatically reduces cycle time.
  • Touch panel features simple, interactive operation.
  • Communications functionality and external I/O support allow automatic testing on production lines.

Product Details

HIOKI ST5540 Leak Current HiTester

Leak current measurement, an essential part of electrical safety (Medical-use Electrical Devices/For Standard- and Regulation-compliant)

IEC 60601-1: (2005) 3rd Edition-compliant. The ST5540 complies with JIS, IEC, and UL standards governing medical- and general-use electrical devices. Uninterrupted polarity switching function dramatically reduces cycle time. Support for rated currents up to 20 A gives the instrument more than adequate capability for testing products designed to comply with new standards. Touch panel features simple, interactive operation.
Communications functionality and external I/O support allow automatic testing on production lines.

Note: Always use an isolation transformer when measuring leak current for medical-use electrical devices. The ST5540 does not include an isolation transformer. When measuring medical-use electrical devices, use a step-up isolation transformer or similar component operating at 110% of the rated supply voltage as the power supply for the device under test.

Basic specifications (Accuracy guaranteed for 1 year)

Measurement methods : Measurement of voltage drop across body simulated resistance points
Calculation and display of current values True rms measurement
Measurement unit floats relative to instrument ground.

Measurement modes : Leak current measurement, voltage measurement, safety conductor current measurement

Standard compliance : Medical electrical equipment: IEC 60601-1 (1988) + A2:1995, (2005, 3rd Edition), JIS T0601-1:1999
Electrical Appliances and Materials Safety Act
Measurement of touch current and protective conductor current: IEC 60990 (1999)
Electrical equipment for measurement, control, and laboratory use: IEC 61010-1 (2001)
Information technology equipment: IEC 60950-1 (2005)
Household and similar appliances: IEC 60335-1 (2001) + A1:2004 + A2:2006
Audio, video and similar electronic apparatus: IEC 60065 (2001) + A1:2005
Personnel Protection Systems for EV: UL-2231-1 (2002), UL-2231-2 (2002)
UL: UL-1492 (1996)
Electrical equipment for measurement, control, and laboratory use; current measurement circuits in damp conditions: IEC 61010-1 (2001)

Leak current measurement : Ground leak current, 3 types of contact current, 7 types of patient leak current, patient measurement current, 4 types of total patient leak current, free current measurement, 3 types of enclosure leak current

Measurement current : DC, AC (true rms, 0.1 Hz to 1 MHz), AC+DC (true rms, 0.1 Hz to 1 MHz), AC peak (15 Hz to 1 MHz)

Measurement ranges : DC / AC / AC+DC mode:50 µA/500 µA/5 mA/50 mA
AC peak mode:500 µA/1 mA/10 mA/75 mA

Measurement accuracy : (current measurement) : DC measurement: ±2.0% rdg. ±6 dgt. (typ.)
AC / AC+DC measurement: ±2.0% rdg. ±6 dgt. (15 Hz to 100 kHz, typ.)
AC peak measurement: ±2.0% rdg. ±6 dgt. (15 Hz to 10 kHz, typ.)

Interfaces : External I/O, medical device relay output, USB 1.1 (communications), RS-232C

Functionality : 110% voltage application, automatic test, data storage for 100 target devices, clock, data backup, printed output (optional), etc.

HiTESTER power supply : 100/120/220/240 V AC (specify at time of order), 50/60 Hz, 30 VA rated power

Target device power supply input : 100 to 240 V AC, 50/60 Hz Rated current input from terminal block: 20 A

Target device power supply output : Output from terminal block: 20 A Output from outlet: 15 A

Dimensions, mass : 320 mm (12.60 in) W × 110 mm (4.33 in) H × 253 mm (9.96 in) D, 4.5 kg (158.7 oz)

Supplied accessories : TEST LEAD L2200 (for ST5540, Red ×2, Black ×1) ×1 set, ENCLOSURE PROBE 9195 ×1, Instruction manual ×1, Users manual ×1, Power cord ×3, Spare fuse for measurement line ×1

Product Description

HIOKI ST5540 Leak Current HiTester consists of measurement networks that simulate the human body and a high-frequency volt meter, and is also capable of switching power supply polarities and making measurements under simulated single fault conditions in equipment under test.

HIOKI ST5541 Leak Current HiTester Leak current measurement, an essential part of electrical safety (For Standard- and Regulation-compliant) LEAK CURRENT HiTESTER ST5541: A low-cost so...
WUHAN HUAYING HYBL 106 Zinc Oxide Arrester Leakage Current Analyzer General introduction HYBL zinc oxide lightning arrester tester uses the advanced microcomputer as the control unit, ...